Cell sorter is used to sort single cells up to 99% purity based on the fluorescent markers using a novel automated sorting setup at droplet frequency of 37-43 kHz. This system is equipped with a sorting system that maintaining a constant stream velocity during runs (sorting are more gently and accurately at high speed).
What kind of research is applicable / suitable for this equipment?
- Size and counts of marine bacteria and other planktons
- Routine monitoring of aquaria waters
- Single cell cultivation, enrichment of mixed particles; PCR and qPCR on sorted cells
Any chargeable costs for equipment usage?
Yes, please approach Lab Manager or search the handbook (Section A6 – Schedule of rates for specialist facilities) for more details.
Technical specifications
Nozzle size (µm) | 100µm (cell size applicable: 0.5 – 80µm) |
Lasers (nm) | Dual laser: 488, 640 nm, 100mW |
Detection | Forward scatter (FSC) with PMT
Side scatter (SSC) with PMT 4 fluorescence detectors with PMT Minimum resolution: 0.5µm |
Droplet frequency | 37 – 43 kHz |
Sensitivity | < 125 MESF for FITC and PE |
Sorting type | Jet-in-air high-performance sorting |
Sorting purity | > 99% |
Temperature range | Sample and collection temperature control system: 4- 37C. |
Example fluorescence | EGFP, PI, Cy5, PE-Cy7 |
Sorting direction | 2-way cell sorting (sort 2 different defined populations at the same time) |
Sorting collection | Up to 5 x 5 ml sample tubes each direction
Up to 5 x 1.5 ml tubes each direction Microscope slides 8-well strip each direction |
Biosafety | Housed in Biosafety System Class I hood |