The SMARTCHEM 450 analyzer uses direct read photometry, utilizes a true discrete and direct read measurement system. It has been widely used in environmental, research laboratories and industrial.
What kind of research is applicable / suitable for this equipment?
- Ideal for large volume of routine analysis, enzymatic or colorimetric work
- Measurement of parameters in all type of water including wastewater, soil & plant, chemical and food
- Relationship between growth response to nutrient supply in plants
Any chargeable costs for equipment usage?
Yes, please approach Lab Manager or search the handbook (Section A6 – Schedule of rates for specialist facilities) for more details.
Technical specifications
Loading capacity | Sample vial: up to 72 slots,
Reagent: up to 32x 50ml reagent bottles or 64x 25ml bottles |
Sampling system | 1 sampling arm for all sampling operation
Automatic probe washing Sample volume: 2.0 – 288 µl (1 µl incr.) Reagent volume: 2.0 – 288 µl (1 µl incr.) Capable of sample pre-dilution, rerun with / without dilution |
Reaction plate | 80 semi disposable cuvettes
Automatic cuvette washing Continuous monitoring of cuvette quality Incubation at 37°C Separate waste lines to minimize hazardous waste |
Dilutor | Automatic pre- and post- dilution |
Max tests / hr | Up to 450 tests (for ISE module) |
Optic system | Source: Halogen 6V/10W; Wavelength 340 to 880nm (9-on board filters); Optical path length: 10mm |
Linear range / Resolution | 0.001 – 4.200 Abs / 0.0001 Abs |
Assay type | End point, kinetic, differential, bi-chromatic assays, fully programmable |
Reading system | Direct photometry |
Minimum reaction volume | 300 µl |
Modules available | ISE module (for pH, redox and conductivity);
Nitrate module with Cadmium reduction |
Example parameters | Alkalinity
Ammonia Chloride Chlorine Chromium Cyanide Fluoride Hardness Iron Manganese Nitrate Nitrite Phenol Phosphate Silicate Sulfate Total Kjeldahl Nitrogen Total Kjeldahl Phosphorous Turbidity Urea TN TP |
Example sample type | Surface water
Ground water Drinking water Waste water Seawater Soils Plants Feed Fertilizers (NOTE: Some parameters require an external pre-treatment of the sample. E.g. digestion / distillation). |